• HVHIPOT GDJS-15A Insulation Tester for Helmet
  • HVHIPOT GDJS-15A Insulation Tester for Helmet
  • HVHIPOT GDJS-15A Insulation Tester for Helmet
  • HVHIPOT GDJS-15A Insulation Tester for Helmet
  • HVHIPOT GDJS-15A Insulation Tester for Helmet

HVHIPOT GDJS-15A Insulation Tester for Helmet

After-sales Service: Yes
Warranty: 1 year
Application: Safety tools
Rated capacity: 5kVA, can be customized
Rated voltage: 50kV, can be customized
Transport Package: Plywood Case
Customization:

360° Virtual Tour

Diamond Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company

Basic Info.

Model NO.
GDJS-15A
Specification
ISO9001
Trademark
HV HIPOT
Origin
Wuhan, China
HS Code
9032339000
Production Capacity
2000

Product Description

Customized GDJS-15A Safety Helmet Test Set 5kVA/50kV Hipot Test Equipment

 

 

I. General Information

 

GDJS-15A Safety Helmet Test Set is special equipment for insulated helmet. The test procedure is simple and personnel security is guaranteed. Parameters such as leakage current, high voltage and withstand voltage curve can be obtained reliably.

The equipment is composed of power frequency voltage withstanding test device, special test electrode for helmet and leakage HV ammeter. It has the advantages of simple wiring and detection, reasonable and reliable layout, convenient movement, flexible dis-assembly and assembly, etc.

 

 

II. Features

7inch color LCD screen, with PLC automation control.

HV Voltage, LV Current, LV voltage, LV current measured at the same time, with high precision sensor and high performance collection chip.

Automatic mode and manual mode for option.

Display HV voltage, LV voltage, LV current, time and withstand voltage result.

Over-voltage and over-current protection. Output voltage, HV current upper limit, LV current upper limit and timing can be set.

Zero-starting function. Test can be proceed after the voltage is back to zero.

Store up to 8000pcs data in 3months. It has USB port to upload data to USB disk.

When setting voltage is arrived, it will timing automatically. After timing, voltage will be back to zero.

When the LV current is over setting value, the voltage will be cut off automatically and it will be back to zero. After the flash-over protection value is exceeded, the automatic step-down zero protection is performed.

Software and hardware anti-interference design, a variety of anti-interference means, to adapt to harsh electromagnetic environment.

Automatic error diagnosis, easy to find and solve problems, simple equipment maintenance.

 

 

Other features

 

Arbitrary combination of high voltage power source, test electrode and leakage ammeter, wiring operation is convenient and flexible.

HV source is oil immersed type testing transformer.

Control unit and HV unit are separate, to ensure safety.

If there is breakdown or over limits, the HV output will be cut off automatically and alarm.

 

III. Technical Specifications

 

A. GDYD-55A Control Unit

 

Rated capacity: 5kVA

Input voltage: AC220V±10%, 60Hz±1

Output voltage: 0-50kVAC

Voltage accuracy: 2.5% (F.S)

Low voltage current: 0-25A

Leakage current: 0-30.0mA

Leakage current resolution: 0.1mA

Current accuracy: 2.0% (F.S)

Timing range: 0-99.9H

Environmental temperature: -20ºC to 50ºC

Environment humidity: 85% RH (no condensing)

 

B. Testing transformer (HV unit) specifications

 

Input voltage: 0-200V

Input current: AC 0-25A

Output voltage: AC 0-50kV

Output current: AC 100mA

Capacity: 5kVA

HV side/LV side turns ratio: 250:1

Measuring voltage: AC 0-100V

    Accuracy: 1.5% of the reading±1 digit (F.S)

No-load current: <10%

Impedance voltage: <10%

Weight: about 45kg

Dimension: 350mm(L)* 250mm(W)* 640mm(H)

 

C. GDM Helmet Electrode

 

Working mode: continuously working

Dimension: 830mm *570mm *1340mm

Weight: 67kg

Max. Withstand voltage: 50kV

Electrode gap: 400mm 500mm, 600mm can be adjustable arbitrarily.

 

D. Working conditions

 

Working place: No gas, steam, chemical dust and explosive and corrosive medium which will influence insulation.

Environment temperature: -20--50ºC

Environment humidity: 85% RH (no condensing)

 