Product Description
GDSF-411WPD 3-in-1 SF6 Gas Analyzer is combined of SF6 humidity (dew point) test, SF6 purity test and SF6 gas decomposition product test. It integrates the functions that can only be achieved with three instruments in one instrument. One-time on-site measurement can complete the detection of all indexes, which greatly saves the gas in the equipment, at the same time reduces the workload of users and improves work efficiency.
Features
1. Zero-point automatic calibration, full curve tracking correction.
2. Protected gas chamber with patented technology, fast measurement speed. l
3. Easy to operate and easy to carry.
4. Good repeatability and fast response.
5. With temperature compensation function
6. AC and DC dual-use, built-in rechargeable lithium battery, automatic switching, overcharge and over discharge protection.
7. Automatic slope calibration.
8. Auto pump back function.
9. Big true-color touch screen, easy to operate.
10. Advanced probe protection.
11. Anti-pollution and anti-interference.
12. High sensitivity and good stability.
Technical parameters
Pressure Range:
Allowable pressure range: intake 0.13 - 1 MPa (absolute pressure)
Display mode:
TFT touch screen 10.1" (resolution 800 x 480)
SF6 Humidity
- Measurement range: dew point -80ºC--+20 or -60ºC--+60ºC(support ppmv)
- Accuracy: ±2ºC(when the dew point temperature is below 0 ºC, the sensor output is the frost point)
- Response time: 63% [90%]
+20→-20ºC Td 5s[45s]
-20→-60ºC Td 10s[240s]
- Resolution: 0.01ºC or 0.01ppm
- Repeat-ability: ± 0.5 ºC
SF6 Purity
- Accuracy: ±0.5%FS
- Repeat-ability: ±0.5%
Decomposition product
- Measuring range:
SO2: 0~1000PPmv----±5%
H2S: 0~100PPmv----±5%
CO: 0~500PPmv----±5%
HF: 0~50PPmv----±5%
- Resolution: 0.01PPmv
- Probe protection: Stainless steel sintered filter
- Power supply: 110~220VAC, AC/DC use
- Lithium battery: AC/DC use with automatic switching, over-charge and over-discharge protection.
- Storage temperature: -40--+70ºC
- Use conditions: Temperature: -35ºC~+60ºC;
Pressure: 0-10bar,
Sampling gas flow rate: not affected
- Other configuration: electronic mass flow meter
- Dimension: 580*420*290mm
- Weight: about 15.5kg.
SF6 Humidity:
Measuring range |
Dew point -80 ~ +20 ºC(support ppmv, etc. ) |
Dew point accuracy |
+-2degree (when the dew point temperature is below 0 degree, the sensor output for the frost point). |
Repeatability |
+-0.5 degree |
SF6 Purity:
Measuring range: |
0 ~ 100% SF6 |
Accuracy and repeatability |
+-0.5%, nothing to do with the flow |